Roll over image to zoom in
Ventilator NEUVM12
$12,990.00$14,700.00 (-12%)
Specifications
Tidal Volume adjustment: 0-1500ml
Tidal Volume: 50~1500ml, error of displayed value: 100 ml and lower than ±20 mL, others ±20 %
Maximum ventilation capacity per minute: ≥18L/m,
Oxygen Concentration: 21%-100%,±15% error allowed
Ventilation Conformance: ≤ 30 mL/kPa
IPPV range: 0~99 times/m, ±15% error allowed.
I/E Ratio: 4:1、3:1.、2:1.、1:1.、1:1.5、1:2.、1:2.5、1:3.、1:4, ±15% error allowed
SIMV: 1-20 times/m, ±15% error allowed.
Maximum Safety Pressure: ≤6.0kPa
Control Pressure: 0.3kPa ~ 4 – PEEP
PTR sensitivity: PEEP-1 cmH2O~PEEP—9 cmH2O adjustable, ±1 cmH2O error allowed.
End-inhale continuous time: 0 ~ 50% adjustable, that is 0.1s ~ 20 s.
PEEP: 0.1 kPa~1.0 kPa
flow trigger sensitivity: Level 3 and Level 8.
Pressure Restriction: 1~6kPa, ±20 % error allowed.
Sigh( deep breath): deep breath in 100 times breath, 1-8 times adjustable( after 10-100 times’
control breath, there will be 1 time deep breath), the inspiration time is 1.5 times to the setup time.
Sustained Operation: 24 hours for main power supply, no less than 30 minutes for built-in
battery application only.