Ventilator Machine NEUVM09

$4,490.00$4,990.00 (-10%)

In stock

SKU: NEUVM09 Category:

Specifications

1.Basic Functions
——Sigh (deep breath);
2.Ventilation Modes
——SIPPV
——IPPV
——IMV
——SIMV
Technical Data
—Tidal volume range: no less than 50 to 1200ml, permissible deviation: ±20 %.
——Maximum minute ventilation: ≥ 18 L/min, permissible deviation: ±20 %.
——Oxygen concentration of output gas: <45 %.
——Controlled ventilation (IPPV) frequency range: 6 ~ 60times/min, permissible deviation: ±15 %.
——I:E ratio: 1:1.5, 1:2.0, 1:2.5 and 1:3.0, permissible deviation: ±15%.
——Maximum safety pressure: ≤6.0 KPa
——Settable maximum operating pressure: 5~6.0 KPa (gas source pressure ranging from 280kPa to 600kPa).
——Oxygen consumption: the variation in the gas pressure in the cylinder should be less than or equal to
1.5MPa/h when the ventilator operates on a 12250KPa / 40L medical oxygen cylinder continuously for one hour.
——Inspiration triggering pressure range: -0.4 ~ 1.0 KPa, permissible deviation: ±0.15 KPa
——Time to switch between controlled and assisted ventilation modes: 6s, permissible deviation: +1 s, -2 s.
——Intermittent mandatory ventilation (IMV) frequency range: 1 to 12 times/min, permissible deviation: ±15%.
——PEEP range: no less than 0.1 to 1.0kPa.
——Sigh (deep breath): the inspiration time should be no less than 1.5 times of the original setting.
——Pressure limit range: 1.0~6.0kPa, permissible deviation: ±20 %
——The presentation of the spontaneous breathing frequency, overall respiratory rate and ventilation capacity is
refreshed once every minute.
——Continuous operation duration: the ventilator can operate continuously on a 24-hour basis on an AC utility main.
——Main unit net weight: 28kg, dimension (L×W×H): 410×300×1100 (mm).